Skip to content
Home » Products » OEM Peristaltic Pump » B200K1/K2/K3KD10 Peristaltic Pump

B200K1/K2/K3KD10 Peristaltic Pump

Introduction

•This product series includes three models: B200K1KD10 / B200K2KD10 / B200K3KD10;
•Control mode: analog speed regulation (4-20mA/0-5V/0-10V optional);
•External level signal control start/stop/direction function;
•With running status output function;
•It can adapt to various tubes
•Small and compact with an attractive and aesthetically pleasing design
•Low running noise;
•Ideal for continuous operation at low speeds.

Parameters

Motor types: 42 stepper motor
Power supply voltage: DC 12 -24V
Wattage: <20W
Speed range: ≤200rpm
Control mode: Analog (4-20mA/0-5V/0-10V)
Suitable tubes: Silicone tube / Pharmed tube
Start-stop method: External Level Signal Control (DC 12-24V)
Direction switch method: External Level Signal Control (DC 12-24V)
Running direction: Clockwise/Counterclockwise rotation
Material of pump head shell: PVDF
Roller material: PVDF
Material of the roller shaft: S304
Material of the mounting plate: S304
Noise: ≤60dB (testing environment noise ≤40dB, horizontal distance between test product and noise meter is 1 meter)
Driver weight: 600g
Dimensions: (L*W*H) 115*60*73mm
Working environment: Temperature 0- 40°C, , Relative humidity< 85% RH
Storage environment: In a clean and well-ventilated environment with ambient temperatures ranging from -40 to +50°C,and relative humidity not exceeding 95%, the air must not contain corrosive, flammable gases, oil mist, or dust.

Tube Model and Flow Reference Table

Tube Material Tube Size mL/rpm 10rpm 30rpm 50rpm 80rpm 100rpm 200rpm
Silicone 0.5*092 0.018 0.18 0.54 0.90 1.4 1.8 3.6
1*0.92 0.058 0.58 1.7 2.9 4.6 5.8 12
2*0.92 0.18 1.8 5.4 9.0 13 18 36
2.4*0.92 0.29 2.9 8.7 15 23 29 58
3*0.92 0.40 4.0 12 20 32 40 80
Pharmed 0.25*0.9 0.0040 0.040 0.12 0.20 0.32 0.40 0.80
0.76*0.85 0.02 0.020 0.60 1.0 1.6 2.0 4.0
1.3*0.85 0.007 0.070 2.10 3.5 5.6 7.0 14
1.52*0.85 0.10 1.0 3.0 5.0 8.0 10 20
1.85*0.85 0.16 1.6 4.7 7.8 12 16 31
2.79*0.84 0.38 3.8 11 19 30 38 76
3.2*0.8 0.48 4.8 14 24 18 48 96

•The above flow data were all tested using a Lead Fluid silicone tube to pump pure water under laboratory conditions with normal temperature and pressure. This data is for reference only.
•Due to pressure in actual use , temperature, medium characteristics, tube material and other specific factors, the specific situation needs to consult our engineers.

Head Pump Structure

Component name and function:
• Pump head pressure block: By pressing the disassembly pressing area on both sides of the pump head pressure block inward and lifting
upwards, the pump head pressure block can be disassembled; conversely, by pressing downwards after installing the conduit, the pump
head pressure block can be installed.
• Tube slot: The inlet and outlet tubes are fixed in place through the tube slot.
• The stepper motor: The rotation of the stepper motor drives the rotation of the roller.
• The drive control unit: manages the stepper motor’s speed, direction, and operating duration settings.
• Mounting plate: It attaches to the customer’s equipment using four through holes on the mounting plate.

Figure 1 Products Structure

Tube Installation

• Press the regions marked by arrows 1 and 2 in Figure 4 with your hand, then lift in the direction of arrow 3 to detach the pump head pressure block.
• Move Arrows 4 and 5 in the direction opposite to what is indicated, and detach the tube slot;
• Position the center of the hose above the pump’s roller mechanism ( this should be situated below the pump head pressing block before it is removed ) ;
• Insert the two disassembled tube slots into the inner sides of the hose cover/tube connector, and then insert them into the pump head in the direction indicated by the arrows in the figure below.
• Press and hold the pressing areas on the sides of the pump head compression block (indicated by arrow 1 and arrow 2). Push the compression block downward in the direction opposite to arrow 3 until the block is securely placed on the pump head.

Figure2 Pump Head Installation Diagram

Dimension(mm)