Function and Feature
• Assembly 1 plastic syringes or gas tight syringes.
• Multiple working mode options.
• Color LCD touch screen, convenient operation.
• Support screen lock, key mute operation.
• Button with indicator light,clearly working state.
• support for a variety of syringes, can be customized syringes.
• High accuracy control.
• Syringe protection and traffic jam alarm.
• RS485 communication,support the MODBUS protocol.
• External signal contol start-stop and direction.
• Wide range power input.
• All-metal anti-static, anti-corrosion spraying process housing.
Technical Parameters
• Work mode: Infuse,withdraw,infuse /withdraw,withdraw/infuse,continuous mode
• Number of channel: 1
• Stroke of pump: 140mm
• Advance per microstep: 0.156μm/μstep
• Linear speed: 1μm/min~150mm/min
• Accuracy: Error<±0.35%(>30% of pump stroke)
• Rated linear force: >20kgf
• Thrust regulation: 1~100% arbitrary adjustable
• Syringe size: 10μL~140mL
• Syringe selection: Built-in main manufacturers, the main model syringe to choose.Can use custom syringe, direct input syringe size and diameter
• Flow rate: 0.184nL/min~173.718mL/min
• Flow calibration: Through the calibration procedure to obtain more accurate fluid volume
• Display: 4.3’color LCD screen display and transfusion volume, residual fluid volume,flow, operation direction ,syringe specification, etc. Animation display operation
• Power-off memory: Storing the running parameters automatically
• Other function: Pause and stop, the sound tip, the lock parameters,fast forward and retreat quickly, display brightness adjustment
• State signal output: One start-stop status,one direction status
• Control signal input: One trigger start,one trigger stop
• Communication interface: RS485,MODBUS Protocol
• Dimension mm: 290×210×180mm
• Weight: 4.4kg
• Power supply: AC100~240V,50/60Hz
• Temperature: Temperature 0 ~ 40℃, relative humidity<80%
Reference Table of Syringe Specifications and Flow Rate
Syringe size | Syringe ID(mm) | Min. flow rate (nL/min) | Max .flow rate(μL/min) |
10μL |
0.485 |
0.184 |
0.027 |
25μL |
0.729 |
0.417 |
0.0626 |
50μL |
1.03 |
0.833 |
0.125 |
100μL |
1.457 |
1.667 |
0.250 |
250μL |
2.304 |
4.169 |
0.625 |
500μL |
3.256 |
8.326 |
1.248 |
1mL |
4.699 |
17.342 |
2.601 |
5mL |
11.989 |
112.890 |
16.933 |
10mL |
14.427 |
163.469 |
24.520 |
20mL |
19.05 |
285.027 |
42.754 |
30mL |
21.59 |
366.090 |
54.913 |
60mL |
26.594 |
555.459 |
83.318 |
100mL |
35.7 |
1000.982 |
150.147 |
140mL |
38.4 |
1158.117 |
173.718 |
Above flow parameters are obtained by using silicone tube to transfer pure water under normal temperature and pressure, in actually it is effected by specific factors such as pressure, medium, etc. Above for reference only.
Dimension (mm)
Statement:The final explanation right of above information belongs to Lead Fluid.