Introduction
•Control mode utilizes RS485 communication (Modbus RTU);
•With running status output function;
•It can adapt to various tubes
•Small and compact with an attractive and aesthetically pleasing design
•Low running noise;
•Ideal for continuous operation at low speeds.
Parameters
Motor types: 42 stepper motor
Power supply voltage: DC12-24V
Wattage: <20W
Speed range: ≤150rpm
Control mode: RS485 (Modbus RTU)
Baud: 9600
Check bit: Parity check/No parity check (configurable via DIP switch)
Data bit: 8
Stop bit: 1
Start-stop method: External Pulse Signal Control (DC 12-24V) / RS485 Communication
Direction switch method: External Pulse Signal Control (DC 12-24V) / RS485 Communication
Running direction: Clockwise/Counterclockwise rotation
Channel number: 1 channels
Roller number: 4 Rollers
Suitable wall thickness of tubes: 0.85 mm
Suitable inner diameter of tubes: 1.52mm/2.06mm/2.79mm
Tube material: Pharmed
Tube installation method: Tube Connector
Tube-pressing type: Spring adjustable
Pump head pressure block material: PVDF
Material of the pump head end cover: PET
Pump head roller material: PVDF
Pump head life: ≥1000h
Noise: ≤60dB (testing environment noise ≤40dB, horizontal distance between test product and noise meter is 1 meter)
Driver weight: 553g
Dimensions: (L*W*H) 103*80*60(mm)
Working environment: Temperature 0- 40°C, , Relative humidity< 85% RH
Storage environment: In a clean and well-ventilated environment with ambient temperatures ranging from -40 to +50°C, and relative humidity not exceeding 95%, the air must not contain corrosive, flammable gases, oil mist, or dust.
Tube Model and Flow Reference Table
Tube | Tube Size | Flow Range(mL/min) | ||||||
1rpm | 20rpm | 40rpm | 60rpm | 80rpm | 100rpm | 200rpm | ||
Pharmed | 1.52*0.85 | 0.1 | 2 | 4 | 6 | 8 | 10 | 15 |
2.06*0.85 | 0.15 | 3 | 6 | 9 | 12 | 15 | 22.5 | |
2.79*0.85 | 0.27 | 5.4 | 10.8 | 16.2 | 21.6 | 27 | 40.5 |
•It is recommended to operate intermittently for speeds above 100rpm. An example of an intermittent operation mode is: run for 3 seconds, stop for 2 seconds,and repeat the cycle.
•The above flow data were all tested using a Lead Fluid silicone tube to pump pure water under laboratory conditions with normal temperature and pressure. This data is for reference only.
•Due to pressure in actual use , temperature, medium characteristics, tube material and other specific factors, the specific situation needs to consult our engineers.
Head Pump Structure
Component name and function:
1.Pump head pressure block: Squeezing the tube with rollers.
2.Mounting Plate: Connects the motor to the pump head bracket and can also be used to install the product onto other equipment;
3.Motor: Drives the pump head;
4.Drive and Control Unit: Provides power, sets operating modes;
5.Commutation screw: Matched with the pump head pressure block;
6.Tube: Transmission of liquid;
7.Tube connector: Securing and connecting tubes.
Figure 1 Products Structure
Tube Installation
• Step 1: Turn the hand screw counterclockwise to unscrew it, then remove the pressure plate.
• Step Two: First, turn the directional screw down, then lift the pressure block upwards and remove the tube-line and the tube connectors.
Figure 2 Diagram of Disassembly Step 1 Figure 3 Diagram of Disassembly Step 2